Medical Isolation Transformers

Protect every heartbeat — choose compliant, expertly engineered medical transformer.

Medical isolation transformers also known as hospital transformers, safeguard patients and equipment by delivering galvanic isolation that is clean, stable and have low leakage current. Built to meet the strict IEC/SANS 61558-2-15 standards, they eliminate shock risks and interference in critical care areas.

Keeping lives safe where safety is non-negotiable.

Related Standards

  • SANS 61558-1 ~ General safety requirements for power transformers

  • SANS 61558-2-15 ~ Particular requirements for medical isolation transformers

  • SANS 10142-1-2 ~ Wiring of premises — Medical locations

  • IEC 60364-7-710 ~ Requirements for electrical installations in medical locations

  • SANS 60601-1 ~ Medical electrical equipment — Basic safety and essential performance

Aspect
Compulsory Requirement
Spec
RT-Complies
Special requirements
Particular requirements for isolation transformers for the supply of medical locations.
IEC 61558-2-15 / SANS 61558-2-15
Yes
Monitoring
Secondary circuit typically includes an Insulation Monitoring Device (IMD) to detect any loss of isolation.
IEC 60364-7-710 / SANS 10142-1-2
Yes
Earthing Arrangement
forms part of a medical IT system with insulation monitoring.
IEC 60364-7-710 / SANS 10142-1-2
Yes
Leakage Current Limit
Extremely low — ≤ 0.5 mA under normal operation (critical for patient safety).
IEC 61558-2-15 / SANS 61558-2-15; IEC 60601-1 / SANS 60601-1
Yes
Electrostatic Shield (Screen)
The screen is earthed.
IEC 61558-2-15 / SANS 61558-2-15
Yes
Dielectric Strength
Typically 4–5 kV AC
IEC 61558-2-15 / SANS 61558-2-15
Yes
Insulation System
Must provide double or reinforced insulation per safety requirements.
IEC 61558-2-15 / SANS 61558-2-15
Yes
Isolation Type
Galvanic isolation
IEC 61558-2-15 / SANS 61558-2-15
Yes
Energy Transfer Method
Electromagnetic induction
IEC 61558-1 / SANS 61558-1
Yes
Connection
No direct conductive connection between the two windings.
IEC 61558-2-15 / SANS 61558-2-15
Yes
Purpose
To electrically separate (isolate) the medical circuit from the mains to protect patients and operators from electric shock.
IEC 61558-2-15 / SANS 61558-2-15
Yes
Principle
Power is transferred magnetically through the transformer’s core
IEC 61558-1 / SANS 61558-1
Yes

Reach out to us and learn more about this very special category of transformers 